Jim, the slow leak would show in your measurements. If they are within tolerance, then the leak was inconsequential. No need for post-cal.

If the leak was consequential, the two measurements would not be in tolerance, and one would have to re-do the entire process. Again, post-cal only shows further deflation after acceptable measurements, which really makes the post-cal irrelevant.
×
×
×
×